Clinical significance of dynamic pressure-volume curve in neonatal mechanical ventilation
ZHU Yuan-Peng, MA Li-Ya, WU Zhi-Jun, LU Guang-Jin, HAN Yu-Kun
Chinese Journal of Contemporary Pediatrics ›› 2009, Vol. 11 ›› Issue (07) : 525-528.
Clinical significance of dynamic pressure-volume curve in neonatal mechanical ventilation
OBJECTIVE: To study the characteristics and role of dynamic pressure-volume curve (P-V curve) in neonatal mechanical ventilation. METHODS: A dynamic P-V curve was automatically drawn by the Stephanie ventilator. The slope rate of dynamic P-V curve was measured in 25 neonates who received mechanical ventilation 1, 24, 48 and 72 hrs after ventilation and before weaning from ventilation. Minute-ventilation (MV), mean airway pressure (Pmean), and fraction of inspired oxygen (FiO2) were recorded. The patterns of dynamic P-V curve during abnormal ventilation (resistance to ventilator, part or complete airway obstruction, airway leaking and tracheal catheter exodus) were observed. RESULTS: With the improvement of pulmonary disease, the slope rate of P-V curve and MV increased, Pmean and FiO2 decreased, and the P-V curve shifted to the volume axle. The slope rate of curve 48 and 72 hrs after ventilation and before weaning from ventilation (1.05±0.48, 1.10±0.42 and 1.13±0.37 mL/cmH2O respectively) increased significantly compared with that 1 hr after ventilation (0.76±0.53 mL/cmH2O) (P<0.05 or 0.01). Abnormal ventilation led to abnormal appearance of dynamic P-V curve. CONCLUSIONS: The increasing slope rate of dynamic P-V curve and the curve shifting to volume axle in neonatal mechanical ventilation may be associated with the improvement of pulmonary disease. The appearance changes of the curve may be of value in the assessment of abnormal ventilation.[Chin J Contemp Pediatr, 2009, 11 (7):525-528]
Dynamic pressure-volume curve / Slope rate / Abnormal ventilation / Mechanical ventilation / Neonate
[1]崔红玉,张少丹,崔广奇,刘玉慧,程艳蕊.急性呼吸窘迫综合征呼吸指标的相关性分析及早期诊断价值研究[J].中国煤炭工业医学杂志, 2008, 11(4):509.
[2]崔玉涛,樊寻梅.气道图形分析技术监测机械通气的意义[J].实用儿科临床杂志, 1999, 14(6):364.
[3]Smith J, Vanlierde S, Devlieher H, Daniels H, Eqqermont E. Birth weight and early lung compliance as predictors of short-term outcome in premature infants with respiratory distress syndrome[J].S Afr Med J, 1995, 85(11):1157.
[4]黄育丹,瞿尔力,邓秀睿,徐海滨,苏卫东.呼吸衰竭患儿行机械通气时呼吸力学参数变化及意义[J].山东医药, 2008, 48(10):82-83.
[5]江文辉,邓力,温惠虹,黄伟娟.婴儿肺炎呼吸系统顺应性及阻力的被动流速容量测定[J].中国当代儿科杂志, 2002, 4(5):382.
[6]佘守章,刘继云,阮祥才,彭梠宽.围术期呼吸力学连续气道监测的研究[J].中华麻醉学杂志, 1997, 17(7):396.
[7]Kessler V, Newth CJ, Guttmann J. Analysis of nonlinear volume-dependent respiratory system mechanics in pediatric patients[J]. Pediatr Crit Care Med, 2000, 1(2):111-118.
[8]Nikischin W, Gerhardt T, Everett R, Bancalari E. A new method to analyze lung compliance when pressure-volume relationship is nonlinear[J].Am J Respir Crit Care Med, 1998, 158(4):1052-1060.